Developing Angle Correction Methods For The Laboratory

by M. Robert DeJong, RDMS, RDCS, RVT
Society Of Diagnostic Medical Sonographers
Representative to the Board Of Directors

from the Spring 2000 issue
1 2 NEXT>>


As high as 35 percent of the applications for accreditation received by the ICAVL demonstrate improper angle correction techniques, making angle correction issues one of the most common causes for delayed decisions. The purpose of this article is to review proper angle correction techniques, so that your laboratory or department can understand how the reviewers evaluate your images with respect to angle correction. There are different thoughts about how to angle correct, for example, parallel to the vessel wall versus parallel to blood flow. It is not the purpose of this article to dictate which method to employ. However, all members of a laboratory should use the same technique following the laboratory’s written protocol. In addition, all submitted cases should demonstrate uniform techniques that follow the guidelines established by the authors of the diagnostic criteria.

Comparison of (A) correct and (B) incorrect angle adjustment.

Click here for additional images.

The physical principles revolve around the fact that the vast majority of ultrasound machines detect and calculate the variables of the Doppler equation, except for the angle between the sound beam and blood flow that is cos q. The vascular technologist/sonographer must input this information into the ultrasound unit via the angle correction control. When used correctly, the velocity of the blood flow can then be measured. Usually, the cursor is placed parallel to the walls of the vessel being investigated to determine angle q. The ultrasound system software then uses the value of cos q in the Doppler equation to determine the velocity scale used to measure the Doppler spectral signals. Since we are dealing with the cosine of an angle, errors of varying severity are introduced when an incorrect value is used. When angle correction is not properly applied, the calculated velocity measurements may overestimate or underestimate the severity of disease. Utilizing erroneous velocity measurements to arrive at a final diagnosis may place the disease severity in a higher or lower category and thus adversely impact the care the patient receives.

Errors are amplified when angles of greater than 60 degrees are used in the determination of velocities. At Doppler angles of less then 60 degrees, an error in the angle correction will result in a very small error in the velocity measurement. As the angle increases between 60 to 90 degrees, the cosine of q (and therefore the denominator of the Doppler equation) rapidly approaches zero; therefore, small errors in the measurement of q result in increasingly large errors in the calculated velocity. It is imperative that a Doppler angle of 60 degrees or less be utilized whenever possible for Doppler velocity measurements.

The method of angle correction used must be based on the diagnostic criteria utilized by the laboratory. It is important to use the same techniques as described in the criteria and the same range of angles.

A survey of ICAVL accredited laboratories was undertaken and reported by Beatrice Madrazo, MD, RVT, of William Beaumont Hospital in Michigan in 1998. Based on the survey, Dr. Madrazo reported that approximately 72 percent of accredited laboratories were utilizing the diagnostic criteria published by the Eugene Strandness et al from the University of Washington. The guidelines established by the University of Washington for those specific criteria include using a consistent angle of 60 degrees and with the Doppler adjusted parallel to the vessel wall.

For all duplex examinations, regardless of the criteria being used, the Doppler angle must be 60 degrees or less. When angles of greater than 60 degrees are obtained, care must be taken to either reposition the transducer or the patient so that the Doppler angle is 60 degrees or less. Occasionally, due to the patient's anatomy, this may not be possible. Whenever angles greater than 60 degrees are used, this must be noted and care must be taken in using these velocity measurements in determining the severity of disease and in reaching a diagnostic conclusion.

NEXT>>

 

HELPFUL HINT

While the majority of the laboratory staff reading this article are familiar with Doppler physics, it is necessary to provide the following information as both a review and a reference for the examples used.

Frequency shifts, and therefore the velocities of the red blood cells, can be determined by the Doppler equation as follows:

fD = ft - fr = 2 ft v cosine q/c

(fD is the Doppler shift, ft the transmitted frequency, fr the received frequency, cosine (cos) q the angle between the sound beam and blood flow, v the velocities of the red blood cells, and c the average velocity of sound in soft tissue.) Solving for the velocities of the red blood cells, the equation now becomes:

v = fD c/2 ft cosine q

 
ICAELonline ICANLonline ICAMRLonline ICACTLonline
Copyright 1997-2008 ICAVL, 8830 Stanford Boulevard, Suite 306, Columbia, MD 21045. All rights reserved.